Abstract

A high-risk vessel of arteriosclerosis is detected by assessing wall shear stress (WSS), which is calculated from the distribution of velocity in a blood flow. A novel echographic method for measuring WSS, which aims to distinguish a normal vessel from a high-risk vessel, is proposed. To achieve this aim, the measurement error should be less than 28.8%. The proposed method is based on a flow model for the area near a vascular wall under a parallel-flow assumption to avoid the influences of error factors. This was verified by an in vitro experiment in which the WSS of a carotid artery phantom was measured. According to the experimental results, the WSS measured by the proposed method correlated with the ground truth measured by particle image velocimetry; in particular, the correlation coefficient and measurement error between them were respectively 0.70 and 27.4%. The proposed method achieved the target measurement performance.