Abstract

A formation of thrombosis due to hemodynamic conditions after the implantation of stent in patent ductus arteriosus (PDA) will derived the development of re-stenosis. The phenomenon of thrombosis formation is significantly related to the distribution of wall shear stress (WSS) on the arterial wall. Thus, the aims of this study is to investigate the distribution of WSS on the arterial wall after the insertion of stent. Three dimensional model of patent ductus arteriosus inserted with different types of commercial stent are modelled. Computational modelling is used to calculate the distributions of WSS on the arterial stented PDA. The hemodynamic parameters such as high WSS and WSSlow are considered in this study. The result shows that the stented PDA with Type III stent has better hemodynamic performance as compared to others stent. This model has the lowest distributions of WSSlow and also the WSS value more than 20 dyne/cm2. From the observed, the stented PDA with stent Type II showed the highest distributions area of WSS more than 20 dyne/cm2. This situation revealed that the high possibility of atherosclerosis to be developed. However, the highest distribution of WSSlow for stented PDA with stent Type II indicated that high possibility of thrombosis to be formed. In conclusion, the stented PDA model calculated with the lowest distributions of WSSlow and WSS value more than 20dyne/cm2 are considered to be performed well in stent hemodynamic performance as compared to other stents.