Abstract
The Medtronic Freestyle stentless bioprosthetic valve provides excellent hemodynamics and long-term durability. However, several studies have reported mid-term structural valve deterioration unique to cases of Freestyle bioprosthesis implantation. According to these reports, the degeneration of the porcine aortic wall causes the dilatation and disruption of the sinus of Valsalva. We encountered a very rare case of an 8 mm-long tear located just above the inflow end of the Dacron cloth. The sudden rupture of the bioprosthetic aortic wall formed a pseudoaneurysm that compressed and almost occluded the pulmonary artery leading to low cardiac output syndrome and shock.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.