Abstract

VWD diagnosis is challenging requiring multiple VWF activity tests using many individual assays. We have developed an ELISA-based VWF Multiplex Activity Assay (VWF-MAA) to address this concern; however, the ability of the VWF-MAA to discriminate between type 1 VWD, variant VWD, and normal subjects has not been evaluated. To evaluate the VWF-MAA and its ability to differentiate between type 1 VWD, variant VWD and normal subjects in individuals undergoing an initial laboratory evaluation for bleeding. A total of 177 plasma samples from the Zimmerman Program: Comparative Effectiveness in the Diagnosis of VWD were evaluated from 11 centres across the US and Canada. The VWF-MAA was compared to Versiti Blood Research Institute (VBRI) and Local Center (LC) assigned VWD diagnosis. Overall, 129/177 (72.9%) were correctly assigned as normal (non-VWD), type 1, or variant VWD compared to the VBRI assigned diagnosis. VWF-MAA assigned non-VWD accurately in 29/57 (50.9%) samples, and type 1 VWD accurately in 93/110 (84.6%) samples. Considering LC diagnosis where there was agreement with VWF-MAA and not VBRI diagnosis, type 1 VWD was accurate in 105/110 (95.5%) samples. Bland-Altman analysis demonstrated good correlation between laboratory methods. VWD, types 2A, 2B, 1C VWD were also assigned by the VWF-MAA. We demonstrate that the VWF-MAA has utility in differentiating type 1 VWD, variant VWD and normal subjects in individuals undergoing an initial laboratory evaluation for bleeding.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call