Abstract
Accurate localization of bone surfaces remains a challenge hampering adoption of ultrasound guidance in computer-assisted orthopaedic surgery. Local phase image features have recently been proven efficacious for segmenting bone surfaces from ultrasound images, but the quality of the processing depends on numerous filter parameters that are currently set through a trial and error process that is tedious, unintuitive and subject to large inter-user variability. A method is presented for automatically selecting parameters of Log-Gabor filters used to extract bone surfaces from 3D ultrasound volumes that is based on properties estimated directly from the specific image. A 15% and 69% average improvement in bone surface localization accuracy on phantom and clinical data, respectively, is demonstrated compared with empirically-set parameters. These findings imply that Log-Gabor filter parameter optimization is necessary for accurate extraction of bone surfaces from ultrasound data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The International Journal of Medical Robotics and Computer Assisted Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.