Abstract

This paper presents a practical procedure for the determination and precision calibration of the volume ratio and absolute value of the volume of vacuum chamber by using static expansion method. Such systems for calculations of volume ratio consist from two vessels connected via a valve called the expansion valve. For determination of absolute value of the volume we added another volume with simple shape (determined by dimensional measurements). In such systems, under the isothermal conditions, the gas pressure is reduced by the ratio of the small volume to the sum of the first volume (small volume) and second volume. The uncertainty of measurement of the volume ratio and absolute volume is calculated as well. Relative volume calibration is performed with uncertainties below 0.75 %.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.