Abstract
Voltammetric measurements were performed for the transfers of Li+ and Na+ at an oil/water interface facilitated by neutral ionophores including 14-crown-4 derivatives. For the organic solvent, nitrobenzene, o-nitrophenyloctylether (NPOE) and o-nitrophenylphenylether (NPPE) were examined. The interface between water and NPPE containing 20 mM dibenzyl-14-crown-4 was found to serve best for the ion-selective-electrode (ISE) surface for the voltammetric determination of Li+. A test of this voltammetric LiMSE using artificial serum showed that the interference from a large amount (ca, 140 mM) of Na+ in blood can be easily avoided by detecting the Na+ concentration simultaneously with the same electrode.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.