Abstract

ABSTRACTAn optimized voltammetric method is reported for the quantification of ibuprofen in aqueous solution using a carbon paste–multiwalled carbon nanotube composite as the working electrode. The electrode preparation and the experimental conditions were optimized; a sensitivity of 228.08 mA mol−1 L was obtained. The limits of detection and quantification were 9.1 µM and 25.5 µM, respectively, and the linear dynamic range extended from 2.36–242 µM. The repeatability was 5.35% expressed as the relative standard deviation. Using statistical analysis for the lower concentrations, the detection limit was reduced to 4.13 µM. The method was employed for the analysis of ibuprofen tablets.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.