Abstract

AbstractVoltammetric behavior of selected genotoxic nitro compounds (2‐nitrofluorene, 2,7‐dinitrofluorene, 2‐nitro‐9‐fluorenone and 2,7‐dinitro‐9‐fluorenone) has been investigated using direct current voltammetry (DCV), differential pulse voltammetry (DPV) and adsorptive stripping differential pulse voltammetry (AdSDPV) at a mercury meniscus modified silver solid amalgam electrode (m‐AgSAE). The optimum conditions have been found for their determination in a 1 : 1 mixture of methanol and aqueous Britton–Robinson buffer solution, with the limits of quantification, (LOQ), 2, 0.3, 0.5, and 0.5 μmol L−1 (DCV at m‐AgSAE) and 0.2, 0.2, 0.4, and 0.2 μmol L−1 (DPV at m‐AgSAE) for 2‐nitrofluorene, 2,7‐dinitrofluorene, 2‐nitro‐9‐fluorenone, and 2,7‐dinitro‐9‐fluorenone, respectively. An attempt at increasing the sensitivity using AdSDPV at m‐AgSAE was successful only for 2‐nitrofluorene (LOQ=2 nmol L−1). Practical applicability of the newly developed methods was verified on direct determination of the studied compounds in drinking and river waters with LOQs ~ 10−7 mol L−1 for all the studied compounds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.