Abstract

A modified carbon paste electrode was prepared by incorporating multi-wall carbon nanotubes with a ferrocene-based polyamide (FDADO-IPC). A mixture of fine graphite powder with 10 wt % of multi-wall carbon nanotubes was applied to the preparation of the carbon paste (by dispersing in paraffin) that was finally modified with a ferrocene polyamide complex. The electrocatalytic oxidation of captopril (CAP) was investigated on the surface of the FDADO-IPC multi-wall carbon nanotubes modified carbon paste electrode (FDADO-IPC-MCNTPE) using cyclic voltammetry (CV), differential pulse voltammetry (DPV), chronoamperometry (CHA) and chronocoloumetry (CHC). Using the modified electrode, the kinetics of CAP electrooxidation was considerably enhanced by lowering the anodic overpotential through a catalytic fashion. A linear dynamic range of 0.2–200 μM for CAP was obtained in buffered solutions at pH 7.0. The detection limit was 0.08 μM. Differential pulse voltammetry as a simple, rapid, sensitive, and selective method was developed for the determination of CAP in tablet and human urine without any treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.