Abstract

A hexagonal mesoporous silica (HMS) functionalized with a 5-mercapto-1-methyltetrazole derivative was employed to prepare a chemically modified carbon paste electrode for Pb(II) detection in aqueous solution by square wave adsorptive stripping voltammetry. The optimal operating conditions were 5 min preconcentration time at pH 6.5, and 120 s electrolysis time in 0.2 mol L−1 HCl. Under these conditions, the voltammetric signal increased linearly with the preconcentration time in the range 1 to 10 min and with the Pb(II) concentration in the range 1 to 100 µg L−1. The electrode was reproducible and sensitive. Simultaneous determination of Pb, Cd and Cu was also carried out with the electrode. The accuracy of the method was validated by analysing Pb(II) in tap water and groundwater samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.