Abstract
Electrochemical properties of disulfiram, representative of highly significant bioactive compounds, were studied with a cyclic renewable silver amalgam film electrode (Hg(Ag)FE) using square wave cathodic stripping voltammetry (SWCSV). The influence of various factors such as pH, buffer concentration, buffer composition, and SWCSV parameters on current response was investigated. The optimum results in terms of signal shape and intensity were recorded in Britton--Robinson buffer (pH 7.5) at -0.5 V versus Ag/AgCl/3 mol L$^{-1}$ KCl. An elaborated electroanalytical procedure was used to determine disulfiram at the Hg(Ag)FE in the concentration range from 0.05 to 5.00 $\mu $M. Precision, repeatability, and accuracy of the method as well as the influence of possible interferences were ascertained. The detection and quantification limits were 11 nM and 37 nM, respectively. The applicability of the developed method was tested in the determination of disulfiram in the commercial formulation Anticol. Thin-layer chromatography with image processing software was used to validate the accuracy of the method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.