Abstract

An ion polymer–metal composite (IPMC) actuator has unique performance characteristics that were applied in this study for use within the eye. Cataracts are a common eye disease causing clouding of the lens. To treat cataracts, surgeons replace clouded lenses with intraocular lenses (IOLs). However, patients who receive this treatment must still wear reading glasses for tasks requiring close-up vision. We suggest a new voltage-controlled accommodating IOL consisting of an IPMC actuator to change the lens’ focus. We examined the relationship between the displacement performance of an IPMC actuator and the accommodating range of the IOL using in vitro experiments. We show that this system has an accommodating range of approximately 1.15 D under an applied voltage of ±1.2 V. By Lagrange interpolation, we estimate that with an IPMC actuator displacement of 0.14 mm, we can achieve a refractive power of 4 D, which is equivalent to the accommodating range of a 40 year old person.