Abstract

This study was designed to test the removal of beta2-microglobulin (beta2M) in a vitamin E-modified membrane. We investigated in vivo the dialyzer (Excebrane, series EE, 1.8 m2) with respect to hydraulic permeability (Kuf), maximum ultrafiltration rate (UF max), sieving coefficient (Sc), and solute clearances in hemodialysis (HD) and in soft hemodiafiltration (HDF). Kuf was 18.4 ml/h/mmHg, UF max was 75 ml/min, and Sc for beta2M was 0.45. Clearance values at 400 ml/min of Qb in HD were 258 ml/min for urea, 201 ml/min for creatinine, and 135 ml/min for phosphate. In soft HDF, clearances were slightly higher. beta2M clearance was 26 ml/min in HD and 43 ml/min in soft HDF. In conclusion, Excebrane (series EE) procures a soft HDF with an amount of substitution fluid in post dilution mode of over 60 ml/min. Remarkable small solute clearances were obtained when the blood flow was raised to 400 ml/min. A significant reduction of beta2M is demonstrated by HDF.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.