Abstract

Colorimetric detection can be applied to differentiate between positive and negative conditions. It can be coupled with loop-mediated isothermal amplification to diagnose rumen fluke or paramphistome infection, also called colorimetric PAR-LAMP. This study conducted LAMP using three candidate indicator dyes, namely malachite green (MLG), methyl green (MTG), and neutral red (NTR), and the results were observed by the naked eye. The dye concentration was optimized to obtain the most pronounced positive-negative result discrimination. Subsequently, we conducted target sensitivity tests using the DNA of Fischoederius elongatus at different concentrations. To validate the detection accuracy, the result was confirmed by gel electrophoresis. The sensitivity test presented the lowest detectable DNA concentration or limit of detection (LOD), with 1pg for MLG, 0.5ng for MTG, and 50pg for NTR. Different LODs revealed inhibition of LAMP reaction and reduced efficiency of result presentation for colorimetric-based detection, particularly NTR and MTG. For MLG-LAMP, we observed no cross-reaction of non-target DNA and improved reaction with the DNA of Fischoederius cobboldi and Calicophoron sp., with multi-detection. In addition, naked eye observation and agarose gel electrophoresis (AGE) evaluation of the MLG-LAMP results showed a moderate and strong agreement with LAMP-AGE and microscopic examinations. Based on our results, colorimetric PAR-LAMP is a rapid, comfortable, and point-of-care procedure for the diagnosis of paramphistome infection.