Abstract

During flexible ureteroscopy (FURS), surgeons may lose orientation due to intrarenal structural similarities and complex shape of the pyelocaliceal cavity. Decision-making required after initially misjudging stone size will also increase the operative time and risk of severe complications. A intraoperative navigation system based on electromagnetic tracking (EMT) and simultaneous localization and mapping (SLAM) was proposed to track the tip of the ureteroscope and reconstruct a dense intrarenal three-dimensional (3D) map. Furthermore, the contour lines of stones were segmented to measure the size. Our system was evaluated on a kidney phantom, achieving an absolute trajectory accuracy root mean square error (RMSE) of 0.6mm. The median error of the longitudinal and transversal measurements was 0.061 and 0.074mm, respectively. The in vivo experiment also demonstrated the effectiveness. The proposed system worked effectively in tracking and measurement. Further, this system can be extended to other surgical applications involving cavities, branches and intelligent robotic surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.