Abstract

Intuitively and efficiently controlling multiple degrees of freedom is a major hurdle in the field of upper limb prosthetics. A virtual reality myoelectric transhumeral prosthesis simulator has been developed for cost-effectively testing novel control algorithms and devices. The system acquires EMG commands and residual limb kinematics, simulates the prosthesis dynamics, and displays the combined residual limb and prosthesis movements in a virtual reality environment that includes force-based interactions with virtual objects. A virtual Box and Block Test is demonstrated. Three normally-limbed subjects performed the simulated test using a sequential and a synchronous control method. With the sequential method, subjects moved on average 6.7±1.9 blocks in 120 seconds, similar to the number of blocks transhumeral amputees are able to move with their physical prostheses during clinical evaluation. With the synchronous method, subjects moved 6.7±2.2 blocks. The virtual reality prosthesis simulator is thus a promising tool for developing and evaluating control methods, prototyping novel prostheses, and training amputees.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.