Abstract

Variation of virtual point detector (VPD) position inside HPGe detector as a function of source photon energy for the energy range from 26.6 to 1,332 keV was investigated. Although VPD concept was well established for HPGe detectors from 59.5 to 10 MeV, a new attempt was made to obtain VPD positions for photon energies below 59.5 keV. It was found that VPD position shows different functional behavior for the energy ranges 26.6–59.5 keV and 59.5–1,332 keV. The VPD position decreases with increasing energy for 26.6, 31.7, 36.4, and 37.3 keV and increases with the energy until it reaches a plateau. The functional behavior of VPD position for the energy range 26.6–59.5 keV was attributed to the dead layer thickness of the Ge crystal. Monte Carlo simulations were performed to investigate the behavior of VPD position with various dead layer thickness ranging from 100 to 800 μm. It was seen that VPD position increases with increasing energy for 31.7, 59.5, and 80.1 keV more significant at relatively lower energies, but constant for the energies 661–1,332 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.