Abstract

The sensor is a kind of device about electrochemical science, its applications include clinical and environmental analyses, physiology, and process control; therefore, how to accurately detect the signal of the sensor is one of the most important things for analyzing the characteristics of sensor. For potentiometric device, this study relates to a multielectrode measurement system based on the programable software, LabVIEW, forming a virtual instrument (VI). This system is built as a voltage versus time (V–T) framework and a dynamic detection system. We selected two devices, a digital multimeter (HP 34401A) and a homemade VI, synchronously to measure the sources of a direct current (DC) signal and an electrode cell (EC), respectively. The maximum errors between the two devices are 0.639 mV in DC supply and 0.345 mV in EC supply, which specifies that the efficiency of design measurement system is good for detection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.