Abstract

The study was retrospective over 8 years (2010-2017), with review of patient records and analysis of downloads from the implantable cardiac devices to the virtual clinics. A total of 75 patients were set up for virtual clinic follow-up during the study period, 94.5% with a pacemaker and 5.5% an implantable cardioverter defibrillator. The majority (76.8%) had an epicardial system. Data on lead impedance, battery longevity, programmed parameters, detected arrhythmias, percentage pacing and delivered defibrillator therapies were obtainable by download. Lead threshold measurements were obtainable via download in 83.7% of the devices, including epicardial systems. No concerning device issue was missed. In 15% of patients a major issue was detected remotely, including three patients with lead fractures. The virtual clinics resulted in fewer hospital attendances while enhancing monitoring and enabling more frequent device checks. The vast majority (91.4%) of families who responded to a questionnaire were satisfied with the virtual clinic follow-up. Virtual clinics allow safe and effective follow-up of children with pacemakers and implantable cardioverter defibrillators, including those with epicardial systems and are associated with high levels of parent satisfaction.