Abstract

We developed a vibration-free cryocooler system based on a 4 K pulse tube (PT) cryocooler for a cryogenic interferometric gravitational wave detector. In this system, we incorporated a vibration-reduction system that consists of a firm support for the cold head and a rigid cold table thermally linked to the cold stage to reduce both the vibrations of the overall cold head and those of the cold stage. Our cryocooler system reduced the cold-head vibration of the baseline PT cryocooler by over three orders of magnitude, where the vibration level was on the order of \( 10^{ - 9} /f^2 m/\sqrt {Hz}\).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.