Abstract

A data recording and processing system is described which transfers gas-phase intensity data directly to punched paper tape, uses a digital computer for all phases of the data handling, and thus substantially reduces the time for the measurement of a single band. The recorder pen position is continually converted to a digital record by a shaft encoder operating at positions on the spectrometer lead screw that are determined by a Schmidt-Trigger encoder unit. Auxiliary information and data output are recorded by utilizing a series of programs that eventually generate sets of partial differentialp/( partial differentialS(i)). The use of this system is illustrated by considering its application to a determination of the absorption band intensities of trans HFCCFH.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call