Abstract
ABSTRACTThe aim of this study was to provide a description and characterize the performance of a volatility hygroscopicity tandem differential mobility analyzer (VH-TDMA) by calibration measurements. In our investigations, we used the two calibration standards most often used by the TDMA community: ammonium sulfate ((NH4)2SO4) and sodium chloride (NaCl) particles. The hygroscopic growth factors, volatility factors, and the hygroscopic growth factors of the nonvolatile particle core were measured in different relative humidity and thermal denuder temperature conditions. The measured hygroscopic growth and deliquescence relative humidities for ammonium sulfate and sodium chloride particles are in line with theory and results from previous measurements, as are the measurements for volatility and hygroscopicity of the particle core. We summarize the measurement campaigns, where the VH-TDMA has been deployed, and show the instrument is capable of measuring high quality data of atmospheric particle properties i...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.