Abstract

Studies to increase the sensitivity of cadmium determination by atomic absorption spectrometry using continuous hydride generation from a vesicular medium within situ preconcentration in a graphite tube are described. Parameters influencing the efficiency of the continuous generation and transport of cadmium hydride are studied in detail using a Pd-coated graphite platform for in situ preconcentration. Continuous stirring of the sample solutions enhanced, approximately ten-fold, the determination of cadmium. The detection limit (LOD) observed for Cd was 5 ng l–1 when preconcentrating 1.4 ml of sample. Improved LOD can be achieved by preconcentrating larger volumes of sample solution, limited only by a maximum deposition of 0.2 ng Cd on to the platform. The observed relative standard deviation for ten replicate analyses of 0.14 ng of total Cd (1.4 ml of a 100 ng l–1 solution) was excellent (±1.1%). The use of W- and Zr-coated platforms and uncoated graphite tubes for preconcentrating Cd in situ was compared with the Pd-coated platform procedure described. The proposed method has been applied successfully to the determination of ng l–1 levels of this toxic metal in river water and tea infusions.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call