Abstract

It has been found that the amino acid analyser used in this study systematically overestimated plasma leucine at high concentrations. The concentration reported as 2249 μmol/L had a true value of 1430 μmol/L and leucine values reported as >2000 μmol/L were approximately 1500 μmol/L.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.