Abstract

Speckle noise reduction is best tested on a precise speckle contrast measurement bench, which should be able to measure 100% contrast in fully developed speckle as well as the smallest contrast (for example, less than 10%) after its reduction. On such a test bench, we have measured very efficient speckle contrast reduction by temporal averaging using a moving diffuser on a tuning fork, which vibrates at 100 Hz over 60 microm in amplitude, a distance that is three times the surface roughness correlation length of the diffuser.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.