Abstract

Calcium contents of demineralised human cortical bone determined by titrimetric assay and atomic absorption spectrophotometry technique were verified by comparing to neutron activation analysis which has high recovery of more than 90%. Conversion factors determined from the comparison is necessary to correct the calcium content for each technique. Femurs from cadaveric donors were cut into cortical rings and demineralised in 0.5M hydrochloric acid for varying immersion times. Initial calcium content in the cortical bone measured by titration was 4.57%, only 21% of the measurement by neutron activation analysis; while measured by atomic absorption spectrophotometer was 13.4%, only 61% of neutron activation analysis. By comparing more readings with the measurements by neutron activation analysis with 93% recovery, a conversion factor of 4.83 was verified and applied for the readings by titration and 1.45 for atomic absorption spectrophotometer in calculating the correct calcium contents. The residual calcium content started to reduce after the cortical bone was demineralised in hydrochloric acid for 8h and reduced to 13% after 24h. Using the linear relationship, the residual calcium content could be reduced to less than 8% after immersion in hydrochloric acid for 40h. Atomic absorption spectrophotometry technique is the method of choice for calcium content determination as it is more reliable compared to titrimetric assay.