Abstract

Graphene liquid cells (GLCs) present the thinnest possible sample enclosures for liquid phase electron microscopy. However, the actual presence of liquid within a GLC is not always guaranteed. Of key importance is to reliably test the presence of the liquid, which is most frequently water or saline. Here, the commonly used methods for verifying the presence of water were evaluated. It is shown that depending on the type of sample, applying a single criterion does not always conclusively verify the presence of water. Testing liquid filling for a specific GLC sample preparation protocol should thus be considered critically. The most reliable method is direct observation of the water exciton peak using electron energy loss spectroscopy (EELS). But if this method cannot be carried out, water filling of the GLC can be verified from a combination of higher contrast in the image, the presence of bubbles, and an oxygen signal in the EEL spectrum, which can be accomplished at a high electron dose in spot mode. Nanoparticle movement does not always occur in a GLC.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.