Abstract

IntroductionThe aim of this study was to examine whether the Canadian Laboratory Initiative on Paediatric Reference Intervals (CALIPER) could be applied to Croatian children and adolescents.Materials and methodsA total of 295 outpatient healthy children and adolescents of age 1 to 18 were selected using the direct a posteriori sampling method. According to current guidelines, 20 samples were tested for each of a total of 51 reference intervals for ferritin, cortisol, dehydroepiandrosterone sulfate, follicle stimulating hormone, lutein stimulating hormone, prolactin, progesterone, sex hormone binding globulin, thyroid stimulating hormone, total testosterone, total thyroxine and total triiodothyronine. Serum samples were analysed on the Beckman Coulter DxI600 immunoassay analyser by chemiluminescence immunoassay method. A reference interval was adopted if < 10% of the results fall outside CALIPER reference interval range. For analytes in which this criterion is not met in the first set of samples, a new set of 20 samples were collected.ResultsAfter the first set of measurements, 96% of all tested reference intervals were adopted for use. The additional sets of 20 reference subjects were tested for only two reference intervals; follicle stimulating hormone for female aged 1 to 9 years, and irrespective of the gender, sex hormone binding globulin for children aged 8 to 11 years. All results of additional samples were within the specified interval limits.ConclusionsCALIPER reference intervals for ferritin and 11 hormones defined for Beckman Coulter DxI600 immunoassay analyser can be implemented into the Croatian laboratories and clinical practice.

Highlights

  • The aim of this study was to examine whether the Canadian Laboratory Initiative on Paediatric Reference Intervals (CALIPER) could be applied to Croatian children and adolescents

  • The additional sets of 20 reference subjects were tested for only two reference intervals; follicle stimulating hormone for female aged 1 to 9 years, and irrespective of the gender, sex hormone binding globulin for children aged 8 to 11 years

  • CALIPER reference intervals for ferritin and 11 hormones defined for Beckman Coulter DxI600 immunoassay analyser can be implemented into the Croatian laboratories and clinical practice

Read more

Summary

Introduction

The aim of this study was to examine whether the Canadian Laboratory Initiative on Paediatric Reference Intervals (CALIPER) could be applied to Croatian children and adolescents. It is necessary to establish reference intervals of diagnostic laboratory tests on referent, or „healthy“, subjects. Reference values based evaluation is one of the options for accurate interpretation of laboratory test results and clinical differentiation between healthy and diseased individuals [1]. According to the European Directive on in vitro diagnostic medical devices, diagnostic test manufacturers are obliged to provide the information about reference intervals for routine test to their customers [2]. Paediatric specific reference intervals for neonatal population, children and adolescents are very often not declared in diagnostics test assay sheets.

Objectives
Methods
Results
Conclusion