Abstract

A small computer has been interfaced to a linear accelerator used for routine radiotherapy. The objectives are to assess the rate and magnitude of some of the mistakes which may occur, to prevent these where possible and to assist radiographers with record keeping. In this paper we describe our approach to the problem, the apparatus used, the computer program, and the results of several years of operating experience while the system was continually being developed. On one linear accelerator, we estimate that an error has been prevented roughly once in every 500 field parameters set. The most common error is in setting the diaphragms and a detailed analysis is given of possible reasons. Doses in monitor units are recorded automatically and a running total of cumulative dose is maintained; plans for the printing of the dose in rads at selected points are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.