Abstract
The simplified arrangement for constancy checks of direct-radiography X-ray units comprises a dosemeter D1 integrated into a 25 mm Al absorber which is inserted into the filter fitting, and a second dosemeter D2 positioned in the cassette drawer of the Bucky table. The measured values of D1, D2 and the quotient D2/D1 yield defined information about kV, mAs and the equipment attenuation factor. The method is simple, independent from film and developer systems, and sensitive for variations in the setup. We recommend it for adoption into the standard DIN 6868.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.