Abstract

Arterial injury resulting from the interaction of contrast agent (CA) with ultrasound (US) was studied in rabbit auricular arteries and assessed by histopathologic evaluation and s-thrombomodulin concentrations. Three sites on each artery were exposed (2.8 MHz, 5-min exposure duration, 10-Hz pulse repetition frequency, 1.4-μs pulse duration) using one of three in situ peak rarefactional pressures (0.85, 3.9 or 9.5 MPa). Saline, saline/CA, and saline/US infusion groups ( n = 28) did not have histopathologic damage. The saline/CA/US infusion group ( n = 10) at exposure conditions below the FDA mechanical index limit of 1.9 did not have histopathologic damage, whereas the saline/CA/US infusion group ( n = 9) at exposure conditions above the FDA limit did have damage (5 of 9 arteries). Lesions were characteristic of acute coagulative necrosis. Mean s-thrombomodulin concentrations, a marker for endothelial cell injury, were highest in rabbits exposed to US at 0.85 and 3.9 MPa, suggesting that vascular injury may be physiological and not accompanied by irreversible cellular injury. (E-mail: zacharyj@uiuc.edu)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.