Abstract

A percutaneous intra-aortic balloon pump (PIABP) was inserted in 112 patients. Successful placement was achieved in 102 patients, 67 of whom survived long enough to have the device removed. Vascular complications related to the PIABP were noted in 15 patients. Reversal of ischemic signs followed PIABP removal in six patients, but nine required femoral artery exploration for thrombectomy, repair of femoral laceration, or repair of false aneurysm. No patient died as a result of PIABP or correction of associated vascular complications, though five of the patients with vascular problems died of cardiac complications. Nine of ten survivors in this group were asymptomatic, and one had persistent paresthesias six months postoperatively. Most clinically significant vascular complications were due to the technique of balloon removal, and several modifications that were effected recently are expected to decrease the incidence of complications requiring surgical intervention.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.