Abstract

Background: Monitoring of vascular access outflow (VAO) in dialysis is based on the indicator dilution method by ultrasound (UD). The role of arterial needle orientation in VAO measurement is not clear. We compared the impact of the retrograde (RET) versus the antegrade orientation (ANT) in terms of (a) VAO (UD) and (b) dialysis adequacy. Moreover, we compared VAO (UD ANT and RET orientation) with VAO measured by Doppler ultrasound. Methods: 22 patients participated in the study. Inclusion criteria: Dialysis > 6 months with a functioning AVF, no stenosis, no active infection, EF > 45% and informed consent. 4 flow measurements were taken on the same dialysis day (4 consecutive weeks). To account for blood pressure variation, we “corrected” VAO for a mean arterial pressure of 100 mmHg. Doppler VAO was measured just before dialysis. Means were compared by the paired t-test. For correlation and agreement, linear regression and Bland-Altman analysis were performed respectively. Results: Mean VAO (UD) was higher in the (ANT) versus the (RET) orientation: 1286.17 mL/min (SD = 455.78, 95%CI = 1084–1488) versus 1189.96 mL/min (SD = 401.05, 95%CI = 1012–1368) (p = 0.013) with a mean difference of 96.21 mL/min (5.66%). Mean Kt/V (RET orientation) was 1.57 (SD = 0.10, 95%CI = 1.52–1.61) versus 1,55 (SD = 0.10, 95%CI = 1.50–1.60) (ANT) orientation (p = 0.062). Recirculation was always 0%. The mean VAO (Doppler) was 1079.54 mL/min (SD = 356.04, 95%CI = 922–1237), 16% lower than VAO measured by UD with (ANT) orientation (p = 0.009) and 9.3% lower than the VAO in the (RET) orientation (p = 0.113). Linear regression analysis showed that VA flows (ANT versus RET) orientation of the needle correlates well between them (r = 0.93, p < 0.001) but show poor agreement (Bland–Altman analysis). Conclusion: VAO (UD) in the RET orientation was significantly lower than VAO in the ANT orientation and more consistent with VAO assessed by Doppler without influencing dialysis adequacy. Therefore, when using UD for VAO surveillance, the RET orientation should be used.

Highlights

  • Monitoring of vascular access outflow (VAO) in dialysis is based on the indicator dilution method by ultrasound (UD)

  • We investigated a possible correlation between VA flow (ANT)–VA flow

  • The principal findings of our study, are that the retrograde arterial needle orientation results in statistically significant lower VA flows than the antegrade orientation and that VA flows in the RET orientation measured by UD were not statistically significant from VA flows measured by Doppler ultrasound (p = 0.113)

Read more

Summary

Introduction

Monitoring of vascular access outflow (VAO) in dialysis is based on the indicator dilution method by ultrasound (UD). The role of arterial needle orientation in VAO measurement is not clear. We compared the impact of the retrograde (RET) versus the antegrade orientation (ANT). In terms of (a) VAO (UD) and (b) dialysis adequacy. We compared VAO (UD ANT and RET orientation) with VAO measured by Doppler ultrasound. Methods: 22 patients participated in the study. 4 flow measurements were taken on the same dialysis day (4 consecutive weeks). To account for blood pressure variation, we “corrected” VAO for a mean arterial pressure of 100 mmHg. Doppler VAO was measured just before dialysis. Means were compared by the paired t-test. Results: Mean VAO (UD) was higher in the (ANT) versus the (RET)

Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.