Abstract

We have developed a blood flow velocimeter to measure maximum blood flow velocity (MBFV) in carotid, brachial and femoral arteries simultaneously by using a Doppler ultrasound technique. This measurement system is based on a continuous wave Doppler ultrasound method with two semicircular piezoelectric (PZT) transducers, one continuously transmitting ultrasound, and the other continuously receiving the echoes. This report is to investigate variations of MBFV in 3 arteries of 5 healthy normal male volunteers during passive postural changes in supine, 90 degrees head-down-tilt (HDT) and 90 degrees head-up-tilt (HUT) respectively. As a result of these measurements, we could confirm that MBFV in those arteries especially in brachial and femoral are increased by supine to HDT posture and are decreased by supine to HUT posture. But, it's just changed a little in carotid artery. An increasing and a decreasing in MBFV are expected to be an effect of venous pressure and a constant MBFV in carotid artery is caused a cerebral autoregulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.