Abstract

Ventilatory flow rates with either He-O2 (heliox) or N2-O2 (nitrox) mixtures during high frequency jet ventilation (HFJV) were calculated using a mathematical model, and were measured in a physical monoalveolar model of the infant lung. A constant flow was delivered to the model (Vd) drawing with it an entrained flow (Ve). When the inspiratory time (Tl) was long, a back flow (Vr) was generated that increased progressively as the alveolar pressure increased. In order to reduce Vr, the reflux time (Tr), i.e., the time needed for the injected gas to reflux, was measured (Trm) and calculated (Trc) with the model. The Tr depends on both chest-lung compliance and resistance to reflux as well as on the Ve/Vd ratio. For a given setting of the ventilator, values of Trm and Trc were significantly lower with heliox than nitrox (P < 0.01), and the correlation between Trm and Trc were fair with both gases. Tidal volume (VT) was calculated when the Tl was interrupted at Tr. The values of VT were significantly higher with helix than nitrox (P < 0.05). It is concluded that the evaluation of Tr during HFJV may provide useful information for setting the ventilator.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.