A cerium-target X-ray tube is useful in performing cone-beam K-edge angiography because K-series characteristic X-rays from the cerium target are absorbed effectively by iodine-based contrast media. The X-ray generator consists of a main controller and a unit with a high-voltage circuit and a fixed anode X-ray tube. The tube is a 1.0-mm-focus diode with a cerium target and a 0.5-mm-thick beryllium window. The maximum tube voltage and current were 65 kV and 0.4 mA, respectively. Cerium Kα rays were selected out using a barium sulfate filter, and the X-ray intensities without filtering and with a barium sulfate filter were 209 and 16.8 µGy/s, respectively, at 1.0 m from the source with a tube voltage of 60 kV and a current of 0.40 mA. Angiography was performed with an X-ray film using the filter and iodine-based microspheres 15 µm in diameter. In the angiography of nonliving animals, we observed fine blood vessels approximately 100 µm in diameter with high contrasts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.