Abstract

IntroductionThe maximum comfort level (MCL), threshold level (THR) and electrical impedance change in the postoperative period of the cochlear implant for months until they stabilize. The objective of this article is to establish the variation during 5 post-surgical years of impedance, and its relationship with MCL in unilaterally implanted adults. MethodsRetrospective study over 5 years, with 78 adult patients implanted with MED-EL in a tertiary hospital from the year 2000 to 2015. The variation in impedance, MCL and the relationship between them were analyzed in basal (9–12), medial (5–8) and apical electrodes (1–4), performing an inferential ANOVA analysis of repeated measures with comparisons between consecutive times, corrected with Bonferroni criteria. Results33 men (42.3%) and 45 women (57.7%), with a mean age of 52.7±14.6 years. “Stability” was considered the time of follow-up without statistically significant differences between one visit and the next. Changes in impedance in medial electrodes ceased to be statistically significant at 3 months, and in apicals at 6 months, with mean values of 5.84 and 6.43kΩ. MCL stabilized at 2 years in basal and apical electrodes, and at 3 years in medial, with mean values of 24.9, 22.7, and 25.6qu. There was a correlation between MCL and impedance in medium electrodes up to 3 months and in apical ones up to one year. ConclusionsElectrical impedance drops significantly in medial and apical electrodes up to 3 and 6 months. MCL increases significantly up to two years. Impedance is related to MCL up to 6 months.