Abstract

Small changes (±2 °C) in ambient temperature were found to affect the zero absorbance calibration of two dissimilar single-beam spectrophotometers of modular construction. The temperature-sensitive modules of each instrument were located and their thermal responses measured. When all modules were operated at ambient temperature (23–27 °C) and monochromator drift was corrected, the thermal drifts were sufficiently large to cause a significant error in applications in which it is not the normal practice to re-calibrate zero absorbance after each measurement, e.g., atomic-absorption and continuous automated spectrophotometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.