Abstract
The objectives were to determine the surgical modality with the lowest aerosol and droplets generated by commonly used modalities in oropharyngeal and laryngeal surgery. A simulation of oropharyngeal and laryngeal surgery was set up using fresh sheep heads. Four common surgical modalities were utilized: cold steel, electrocautery, coblation, and microdebrider. The resultant aerosol generated was evaluated using two measurement modalities at two key positions in the theater. (1) DustTrak Pro Aerosol Monitor was used to measure the concentration of particles. (2) Fluorescein dye coated on the oropharynx and larynx, and the resultant scatter on paper. Electrocautery and coblation produced statistically significant increases in the concentration of aerosols (p < .001). Microdebrider and cold steel instrumentation produced the least aerosols. No measurable fluorescein droplets were noted for all four modalities. Electrocautery and coblation produced higher concentrations of aerosols. Mitigation factors should be considered with instruments with increased aerosolization. These modalities show low droplet-related infection risk.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.