Abstract
This paper aims to show how the variance reduction technique “Geometry splitting/Russian roulette” improves the statistical error and reduces uncertainties in the determination of the absorbed dose rate in tissue using an extrapolation chamber for beta radiation. The results show that the use of this technique can increase the number of events in the chamber cavity leading to a closer approximation of simulation result with the physical problem. There was a good agreement among the experimental measurements, the certificate of manufacture and the simulation results of the absorbed dose rate values and uncertainties. The absorbed dose rate variation coefficient using the variance reduction technique “Geometry splitting/Russian roulette” was 2.85%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.