Abstract
Multiple points and a spatial mesh based surface source model (MPSMBSS) was generated for 18MV Varian 2100 C/D Clinac phase space file (PSF) and implemented in MCNP code. The generated source model (SM) was benchmarked against PSF and measurements. PDDs and profiles were calculated using the SM and original PSF for different field sizes from 5 × 5 to 20 × 20 cm2. Agreement was within 2% of the maximum dose at 100cm SSD for beam profiles at the depths of 4cm and 15cm with respect to the original PSF. Differences between measured and calculated points were less than 2% of the maximum dose or 2mm distance to agreement (DTA) at 100 cm SSD. Thus it can be concluded that the modified MCNP code can be used for radiotherapy calculations including multiple source model (MSM) and using the source biasing capability of MPSMBSS can increase the simulation speed up to 3600 for field sizes smaller than 5 × 5 cm2.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.