Abstract

We describe the construction and operation of a variable temperature (VT) system for a high field fast magic angle spinning (MAS) probe. The probe is used in NMR investigations of biological macromolecules, where stable setting and continuous measurement of the temperature over periods of several days are required in order to prevent sample overheating and degradation. The VT system described is used at and below room temperature. A vortex tube is used to provide cooling in the temperature range of −20 to 20 °C, while a liquid nitrogen-cooled heat exchanger is used below −20 °C. Using this arrangement, the lowest temperature that is practically achievable is −140 °C. Measurement of the air temperature near the spinning rotor is accomplished using a fiber optic thermometer that utilizes the temperature dependence of the absorption edge of GaAs. The absorption edge of GaAs also has a magnetic field dependence that we have measured and corrected for. This dependence was calibrated at several field strengths using the well-known temperature dependence of the 1H chemical shift difference of the protons in methanol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.