Abstract
We demonstrate a design for a scanning superconducting quantum interference device (SQUID) microscope in which the sample temperature can be varied over a large range. In this design, both sample and SQUID are in the same vacuum space, separated by a few microns. By firmly anchoring the SQUID to a low-temperature bath, the sample temperature can be changed while the SQUID remains superconducting. This allows magnetic imaging at varying sample temperatures with micron-scale spatial resolution and the sensitivity of a low-Tc SQUID. We demonstrate this approach by imaging the temperature dependence of Abrikosov vortices in thin films of the high-temperature superconductor YBa2Cu3O7−δ. We extract the in-plane penetration depth λab(T) in our samples from these measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.