Abstract

A new method for measuring trace concentrations of atmospheric pollutants by infrared diode laser spectroscopy has been devised. This method relies on the increase of the signal as the pressure inside the cell increases, while the frequency of the diode is stabilized on the line, even if it is unresolved. Performances of this method were tested with N2O and with 1,3-butadiene. As an example of application, we measured the butadiene emitted by car exhausts. Sensitivity and rapidity of this method are equivalent to the usual scanning method in which the whole line is described, but this new method benefits from its simplicity and robustness.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.