Abstract

Background The quantification of serum IgG4 is commonly performed during the diagnostic workup of IgG4-related diseases (IgG4-RD). According to recent literature, IgG4 values above 1.35 g/L are characteristic of IgG4-RD and support its diagnosis at initial presentation. The purpose of this study was to evaluate comparability and accuracy of the two main commercially available IgG4 assays (Siemens Healthineers and The Binding Site). Methods Method comparison was performed for IgG and IgG subclasses using a collective of selected samples with elevated serum IgG4. In addition, we assessed the accuracy of both assays using purified polyclonal and monoclonal IgG4 preparations. Results Our data show significant discrepancies between the two IgG subclass assays for the measurement of IgG4 and, to a lesser extent, IgG3. Conclusions The lack of standardization between the two main providers of commercially available IgG4 assays leads to significant inter-assay result discrepancies, which might potentially cause unnecessary clinical workup. We conclude that serum IgG4 assay-specific decision limits, and not an assay-independent single cut-off level for IgG4 (e.g. 1.35 g/L), should be used when assessing patients for IgG4-RD. An internationally recognized, certified reference material for IgG subclasses is urgently needed, and assay manufactures are encouraged to undertake steps toward standardization of measurements of IgG4 and other IgG subclasses.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.