Abstract

Background: Postural stability and plantar pressure parameters can be assessed by baropodometry; nevertheless, they are often affected by low repeatability. The aim of the study was to test the accuracy and repeatability of a novel resistive sensor pressure plate and to establish the most reliable baropodometric parameters. Methods: Accuracy and repeatability of the FM12050 BTS-Bioengineering plate measurements were assessed by using different weights in static conditions across three sessions. Subsequently, 20 healthy subjects were assessed by 30-s stabilometric analysis in bipedal standing with open eyes across four trials in two sessions, morning and afternoon. Results: Pressure plate repeatability in measuring the static weights was very high, and plate measurements were correlated to the scale measurements (Pearson’s coefficient = 0.99). Percentage of load distribution between left and right foot and in rearfoot and forefoot regions showed the largest repeatability (coefficient of variation < 5%) across trials. Eventually, median and percentiles (25–75%) were reported for each parameter. Conclusions: This study helped to assess the accuracy and repeatability of a novel pressure plate in static conditions and to define the most reliable parameters for the assessment of postural stability and foot morphology. The present healthy-subject stabilometric dataset may be used as reference data in the evaluation of pathological populations.

Highlights

  • Upright posture is maintained by tonic muscle contractions stabilizing the positions of body segments against the force of gravity [1]

  • Ethical review and approval were waived for this study due to the nature of this pilot study, which required the recruitment of a small population of healthy participants tested for standard baropodometric parameters

  • Coefficient cient of variation ranged between for the smallest weight, between of variation ranged between 0.014–0.025 (1.4–2.5%) for the smallest weight, between

Read more

Summary

Introduction

Upright posture is maintained by tonic muscle contractions stabilizing the positions of body segments against the force of gravity [1]. Postural stability and plantar pressure parameters can be assessed by baropodometry; they are often affected by low repeatability. The aim of the study was to test the accuracy and repeatability of a novel resistive sensor pressure plate and to establish the most reliable baropodometric parameters. Methods: Accuracy and repeatability of the FM12050 BTSBioengineering plate measurements were assessed by using different weights in static conditions across three sessions. Conclusions: This study helped to assess the accuracy and repeatability of a novel pressure plate in static conditions and to define the most reliable parameters for the assessment of postural stability and foot morphology. The present healthy-subject stabilometric dataset may be used as reference data in the evaluation of pathological populations

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.