Abstract

Degradation at the modular head-neck interface in total hip arthroplasty (THA) is predominately expressed in the form of corrosion and fretting, potentially causing peri-prosthetic failure by adverse reactions to metal debris. This retrieval study aimed to quantify variations in stem taper surface topographies and to assess the influence on the formation of corrosion and/or fretting in titanium alloy stem tapers combined with metal and ceramic heads. Four hip stem designs (Alloclassic, CLS, Bicontact and SL-Plus) were characterized using high-resolution 3D microscopy, and corrosion and fretting were rated using the Goldberg scoring scheme. Quantification of the taper surface topographies revealed a high variability in surface characteristics between threaded stem tapers: Alloclassic and CLS tapers feature deeply threaded trapezoid-shaped profiles with thread heights over 65 µm. The sawtooth-shaped Bicontact and triangular SL-Plus taper are characterized by low thread heights below 14 µm. Significantly lower corrosion and fretting scores were observed in lightly threaded compared to deeply threaded tapers in ceramic head combinations. No significant differences in corrosion or fretting scores with thread height were found in pairings with metal heads. Understanding the relationship between stem taper surface topography and the formation of corrosion and fretting could help to improve the performance of modern THAs and lead to longer-lasting clinical results.

Highlights

  • Degradation at the modular head-neck interface in total hip arthroplasty (THA) is predominately expressed in the form of corrosion and fretting, potentially causing peri-prosthetic failure by adverse reactions to metal debris

  • The high modularity in total hip arthroplasty enables surgeons to choose between different femoral head sizes and materials, femoral neck lengths, stem sizes and taper characteristics to select the components that best fit the needs of the patient

  • The variability in stem taper surface topography between hip stem designs has been identified as a potential cause of degradation by fretting and corrosion

Read more

Summary

Introduction

Degradation at the modular head-neck interface in total hip arthroplasty (THA) is predominately expressed in the form of corrosion and fretting, potentially causing peri-prosthetic failure by adverse reactions to metal debris This retrieval study aimed to quantify variations in stem taper surface topographies and to assess the influence on the formation of corrosion and/or fretting in titanium alloy stem tapers combined with metal and ceramic heads. In the present study we aimed to further quantify the variability in taper surface topographies of hip stems that are commercially available (Alloclassic, CLS, Bicontact and SL-Plus) and investigate the influence of varying surface finishes on the formation of fretting and corrosion in the head-neck interface. The degree of corrosion and fretting is compared between ceramic and metal femoral heads with simultaneous consideration of the surface finish

Objectives
Methods
Results
Conclusion