Abstract
ObjectivesIn this study, we compared left ventricular (LV) segmental strain measurements obtained with different ultrasound machines and post-processing software packages. BackgroundGlobal longitudinal strain (GLS) has proven to be a reproducible and valuable tool in clinical practice. Data about the reproducibility and intervendor differences of segmental strain measurements, however, are missing. MethodsWe included 63 volunteers with cardiac magnetic resonance–proven infarct scar with segmental LV function ranging from normal to severely impaired. Each subject was examined within 2 h by a single expert sonographer with machines from multiple vendors. All 3 apical views were acquired twice to determine the test-retest and the intervendor variability. Segmental longitudinal peak systolic, end-systolic, and post-systolic strain were measured using 7 vendor-specific systems (Hitachi, Tokyo, Japan; Esaote, Florence, Italy; GE Vingmed Ultrasound, Horten, Norway; Philips, Andover, Massachusetts; Samsung, Seoul, South Korea; Siemens, Mountain View, California; and Toshiba, Otawara, Japan) and 2 independent software packages (Epsilon, Ann Arbor, Michigan; and TOMTEC, Unterschleissheim, Germany) and compared among vendors. ResultsImage quality and tracking feasibility differed among vendors (analysis of variance, p < 0.05). The absolute test-retest difference ranged from 2.5% to 4.9% for peak systolic, 2.6% to 5.0% for end-systolic, and 2.5% to 5.0% for post-systolic strain. The average segmental strain values varied significantly between vendors (up to 4.5%). Segmental strain parameters from each vendor correlated well with the mean of all vendors (r2 range 0.58 to 0.81) but showed very different ranges of values. Bias and limits of agreement were up to −4.6 ± 7.5%. ConclusionsIn contrast to GLS, LV segmental longitudinal strain measurements have a higher variability on top of the known intervendor bias. The fidelity of different software to follow segmental function varies considerably. We conclude that single segmental strain values should be used with caution in the clinic. Segmental strain pattern analysis might be a more robust alternative.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.