Abstract
The purpose of this study was to describe a possible mechanism of failure of the Ahmed glaucoma valve (AGV; New World Medical, Rancho Cucamonga, CA). Two AGVS that failed to control intraocular pressure (IOP) were examined. In vivo examination consisted of unroofing the capsule surrounding the implant and injecting fluorescein-stained balanced salt solution into the tube. In one case the entire implant was removed and examined in vitro, and in the other case only the valve portion was removed. No flow in either case extended beyond the valve. In both cases, the membrane valve system was found to be fused and could not be pulled apart. Failure of the AGV may occur due to adhesion of the valve membranes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.